Clean Bench
Clean Bench
General information
BEING provides high-quality, safe Vertical Laminar Clean Benches. BEING’s clean benches suck air from the room or hall space, transfer the air through a HEPA filter using a fan, and then clean the bench area with filtered air.
● Reservation on/off control for fan & UV light
● High quality, quiet fan, adjustable air volume
● Universal electrical outlet
● An additional pre-filter below work surface helps capture large particles and prolongs HEPA filter life
● Working area is equipped with floodlights and ultraviolet sterilizing lamps
● The sliding tempered glass window with self-balancing design, can be stopped any position that customer want
Spec/Model | BCV-1FD | BCV-1F | BCV-2FD | BCV-2F | BCH-1BU | BCH-1CU |
Num. of operator | one people one side | one people two side | two people one side | two people two sides | one people one side | one people one side |
Laminar Flow | Vertical | Horizontal | ||||
Cleanliness | ISO class 5, ISO-14644-1 | |||||
CFU | ≤0.5/Dish*h (φ90mm petri dish) | |||||
Air Velocity | ≥0.3m/s | |||||
Noise | ≤62dB | |||||
Vibration | ≤3μm | ≤3μm | ≤4μm | ≤4μm | ≤3μm | ≤4μm |
Illumination | ≥300Lx | |||||
Power input | AC220V/50Hz | |||||
Peak power | 250VA | 250VA | 500VA | 500VA | 250VA | 500VA |
External Dimensions W x D x H(mm) | 870×690×520 | 870×690×520 | 1360×690×520 | 1360×690×520 | 825×480×600 | 1680×480×600 |
Internal Work Area W x D x H(mm) | 1010×765×1610 | 1010×780×1610 | 1500×765×1610 | 1500×780×1610 | 900×700×1450 | 1760×700×1450 |