Bluepard Line
Bluepard Line
Salt spray corrosion test chamber

Salt spray corrosion test chamber

Salt spray corrosion test chamber

Features

The overall molding, corrosion resistance, shock, easy to clean and no leakage.

Tower spray system, the salt spray is uniformity and the settling volume is adjustable.

PVC or FRP cover could be optional, using PVC cover could clearly see the test samples and spraying situation in working chamber.

Water seal structure and no salt spray overflow.

The circuit control board and other components are fixed the location which is easy to check and maintain.

Control system

P.I.D Microprocessor temperature controller ensures a precise and reliable control, the resolution is ±0.1℃ .

Universal type: With domestic digital controller.

All circuits are equipped with circuit breakers and all heaters are equipped with overheating protection devices, Thedistribution box is made of refractory resin.

Continuous or periodic spray optional and dual over-temperature protection, which is safe and reliable.

Safety Alarm

With electric leakage, overload protection.

With short circuit, shortage of water, over-temperature alarm.


Specifications
Model

LYW-015

LYW-015N

LYW-025

LYW-025N

LYW-075

LYW-075N

Working chamber dimension(W×D×H,mm)600×450×400900×600×5001100×750×500
External dimension(W×D×H,mm)1150×1100×5601400×1200×8501650×950×1300
Temperature rangeRT+5~50℃
Temperature uniformity±2℃
Temperature stability±0.5℃
Electrical requirementAC220V, 50Hz
Power consumption2000W
Salt spraying sedimentation rate1.0~2.0ml/80cm2
Meet standardGB/2423.17-93、GB/T10587-2006、GB/T10125-2012、ISO9227-1990


Accessory
    No Data